Specification builder · Tier B

Package seal-strength tester: seal peel tester (ASTM F88/F88M, EN 868-5) and/or whole-pouch burst and creep tester (ASTM F1140/F1140M, ASTM F2054/F2054M) for sterile-barrier and flexible packaging

A package seal-strength tester measures how well sterile pouches and flexible packages are sealed, either by peeling a narrow strip cut across the seal (seal strength in N per 15 or 25 mm) or by inflating the whole pouch until it bursts or fails under a held pressure. Drug and device testing laboratories, local makers of sterilisation pouches and of packed devices such as syringes and IV sets, and university packaging labs use it for batch release and for ISO 11607 sealing-process validation. Correct grips, strip cutting, tail technique and accredited calibration matter as much as the instrument.

What this equipment will do

  • Sample types: paper/film and nonwoven/film sterilisation pouches and reels, header bags, lidding of trays and blisters, and other heat-sealed films and laminates, cut into 15 mm or 25 mm strips. (with the optional module)
  • Sample types: complete pouches and bags, open-ended (before filling) through a clamp on the open end. (with the optional module)
  • Complete sealed (finished) packages through a closed-package probe. (with the optional module)
  • Seal strength by peel: maximum and average force per strip width, with the tail unsupported, held at 90 degrees or at 180 degrees (ASTM F88/F88M techniques A, B, C), and the 15 mm strip test for sealable pouches and reels (EN 868-5). (with the optional module)
  • Automatic maximum and average seal strength over a set zone, energy to separate, failure mode and batch statistics. (with the optional module)
  • Whole-pouch strength without restraint: burst pressure, creep (set pressure held for a set time) and creep-to-failure (time until failure at a set pressure) (ASTM F1140/F1140M). (with the optional module)
  • Burst testing between restraining plates, which loads the seals more evenly in large pouches (ASTM F2054/F2054M). (with the optional module)
  • Pressure-hold / pressure-decay check of whole packages with pass/fail. (with the optional module)
  • Strong packages, trays and lids needing pressures up to 600 kPa or more. (with the optional module)
  • Strong lidding seals and film tensile checks above 100 N. (with the optional module)
  • Seal-strength data for validating heat sealers and packaging (ISO 11607-1 and ISO 11607-2 operational and performance qualification).
  • Audit trail, user levels and electronic signatures for results used in batch release or regulatory decisions (21 CFR Part 11). (with the optional module)

Not covered: Gross leak detection of whole packages by bubbles under water (ASTM F2096) or seal-channel detection by dye penetration (ASTM F1929) need the package burst / leak tester and a dye-penetration kit.; Container closure integrity by vacuum decay (ASTM F2338) needs the container closure integrity tester (CCIT).; Accelerated ageing of packages before testing (ASTM F1980) needs the environmental test chamber; making the seals needs the sterilisation pouch sealing machine.; Tensile strength, elongation and puncture of the packaging films themselves are best done on the universal testing machine.

7 international manufacturers checkedMarket checked 2026-1011 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 61010-1 Basic electrical safety standard for laboratory and measurement equipment, including the tester and its controller.
  • IEC 61326-1 EMC standard for laboratory electrical equipment; keeps force and pressure readings stable near other equipment and on noisy mains.
  • ASTM F88/F88M-23 The seal strength (peel) test of flexible barrier materials: strip width, tail-holding techniques A, B and C, maximum and average seal strength.
  • EN 868-5:2018 European requirements and test methods for sealable pouches and reels of porous material and plastic film, including the seal strength test on 15 mm strips.
  • ASTM F1140/F1140M-13(2025) Internal pressurisation failure resistance of unrestrained packages: burst, creep and creep-to-failure tests of whole pouches.
  • ASTM F2054/F2054M-13(2024) Burst testing of flexible package seals using internal air pressurisation within restraining plates.
  • ISO 11607-1:2019 Requirements for sterile barrier systems and packaging systems; seal strength and package integrity tests are used to show conformity.
  • ISO 11607-2:2019 Validation of forming, sealing and assembly processes; peel and burst results are used in sealer operational and performance qualification.
  • ISO 7500-1:2018 Verification and classification of the force-measuring system of tension testing machines, used to calibrate the peel tester's load cells.
  • ISO/IEC 17025:2017 Accreditation standard for testing and calibration labs; needs traceable calibration and controlled records.
  • 21 CFR Part 11 US FDA rule on electronic records and signatures; applies when electronic test results are used for batch release or regulatory decisions.

Common mistakes in tenders for this equipment

  • Copying one maker's brochure (exact speed steps, load cell size, pressure range or software name) so that only one brand complies; state minimum needs with 'not less than' instead.
  • Asking for a high-capacity load cell (500 N to 5 kN) for pouch seals that peel at 1-20 N, where the reading is outside its accurate range; ask for a load cell of not more than 100 N.
  • Specifying ASTM F88 without saying which tail technique (A, B or C) and strip width (15 or 25 mm) the lab must run, so results cannot be compared with the pouch maker's specification.
  • Buying a burst tester without the open-package fixture, closed-package probe or restraining plates that the lab's packages need, so it cannot test the actual products.
  • Combining peel and burst testers in one lot, which only very few makers can supply, instead of two separate lots.
  • Forgetting the compressed-air supply for a burst tester, or using oily, wet workshop air that damages valves and sensors.
  • Accepting a factory certificate instead of an on-site ISO/IEC 17025 accredited force or pressure calibration, which assessors expect.
  • Asking for DRAP medical-device registration or ISO 13485 for a laboratory test instrument; these demands only confuse bidders.