Specification builder · Tier A
Benchtop pH / mV / temperature meter (0.01 or 0.001 pH class) with optional conductivity channel for pharmacopoeial pH and water conductivity
A benchtop pH meter measures the pH, electrode potential (mV) and temperature of liquids with a glass combination electrode; a second channel can add conductivity for purified water and WFI. Pharma QC labs, drug testing labs, hospital labs, water and food labs and universities use it for buffers, mobile phases, dissolution media, liquid products and water. The buying decisions are the resolution class (0.01 or 0.001 pH), whether conductivity is needed, the electrode types for the samples, and whether GMP electronic-record controls are required.
What this equipment will do
- Samples: aqueous solutions, buffers, HPLC mobile phases, dissolution and culture media, oral liquids, injections, eye drops, reagents and water samples.
- Semisolid and surface samples: creams, ointments, gels and agar plates, with a flat-membrane electrode. (with the optional module)
- Small-volume samples (about 1 mL or less in vials and tubes) with a micro electrode. (with the optional module)
- Low-ionic samples such as purified water, WFI and mobile phases with organic solvent, with a low-ionic-strength electrode. (with the optional module)
- pH to 0.01 pH or better by the pharmacopoeial method with at least two bracketing buffers (USP <791>, Ph. Eur. 2.2.3 / BP Appendix V L).
- pH to 0.001 pH with meter accuracy +/-0.003 pH or better, showing electrode drift. (with the optional module)
- mV / ORP (redox potential) readings.
- Sample temperature with automatic temperature compensation of the pH reading.
- Electrode calibration with automatic buffer recognition and slope/offset display.
- Conductivity (with TDS / resistivity / salinity) of water, buffers and solutions (ISO 7888). (with the optional module)
- Pharmaceutical water conductivity test for purified water and WFI, using uncompensated conductivity and a low-constant certified cell (USP <645>, Ph. Eur. 2.2.38). (with the optional module)
- Guided three-stage water conductivity test with pass/fail report. (with the optional module)
- Ion concentration (for example fluoride, chloride, nitrate, ammonium) with ion-selective electrodes. (with the optional module)
- Field and water-loop pH measurement with a portable waterproof meter. (with the optional module)
- GLP records of results and calibrations with date, time and operator or sample ID, printed or exported.
- Electronic records with user login, audit trail and electronic signatures (21 CFR Part 11 / EU GMP Annex 11). (with the optional module)
Not covered: Acid-base, redox or Karl Fischer titrations and water content: these need a potentiometric autotitrator or a Karl Fischer titrator.; Total organic carbon (TOC) of pharmaceutical water (USP <643>) needs a TOC analyser; the conductivity channel does not replace it.; pH of tablets, powders or soils is not measured directly; the sample must first be dissolved or extracted as the monograph directs.; Dissolved oxygen, turbidity, chlorine and heavy metals in water need a DO meter, a turbidimeter, a photometer or AAS/ICP.
Loading the builder...
Standards this specification draws on
- IEC 61010-1 Basic electrical safety standard for laboratory measuring equipment, including the mains adapter.
- IEC 61326-1 EMC requirements for laboratory measuring equipment; high-impedance pH inputs are sensitive to interference.
- USP <791> USP general chapter on pH: meter reading to 0.01 pH or better, calibration with at least two buffers bracketing the sample, and checks of the system.
- Ph. Eur. 2.2.3 European Pharmacopoeia potentiometric pH method (BP Appendix V L), including reference buffers and calibration checks.
- USP <645> Three-stage water conductivity test for purified water and WFI, with uncompensated measurement, known cell constant and meter calibration requirements.
- Ph. Eur. 2.2.38 European Pharmacopoeia conductivity method with instrument calibration and cell-constant verification using certified solutions.
- ISO 10523 Water quality method for pH determination used by environmental and drinking-water labs.
- ISO 7888 Water quality method for electrical conductivity used by environmental and drinking-water labs.
- DIN 19266 Reference buffer solutions for calibrating pH measuring equipment; buffer values recognised by meters are traceable to these and NIST values.
- ISO 17034 Competence of reference-material producers; certified buffers and conductivity standards should carry ISO 17034 (or ISO/IEC 17025) certificates.
- ISO/IEC 17025 Competence standard for laboratories issuing calibration certificates; also the accreditation standard of drug testing labs.
- 21 CFR Part 11 / EU GMP Annex 11 Rules for electronic records, audit trails and signatures where pH results are stored electronically in GMP labs.
Common mistakes in tenders for this equipment
- Copying one brand's exact figures (very large memory counts, a specific screen size, a named number of plug-in modules) so only that model qualifies; use 'not less than' and 'not worse than' limits instead.
- Using proprietary names for intelligent electrodes, electrode-condition checks or compliance software instead of describing the function.
- Asking for a 'USP <645> certified meter' or a '21 CFR Part 11 certified pH meter'; no such certification exists. Ask for uncompensated conductivity, a certified low-constant cell and a Part 11 compliance matrix.
- Forgetting the electrode is a consumable: no spare electrode, no electrolyte, no local price list and no 7-year electrode supply undertaking, so the meter is dead within two years.
- Specifying only 4-7-10 buffers for a lab that measures pH 1-2 (0.1 M HCl dissolution media) or pH 12 (cleaning solutions); bracketing buffers near pH 1.68 and 12.45 are needed.
- Buying a standard 1 /cm conductivity cell for purified water or WFI; readings below 10 uS/cm need a 0.1 /cm (or lower) cell with a certified constant and uncompensated measurement.
- Leaving out certified buffers and conductivity standards with ISO 17034 / ISO/IEC 17025 certificates, then calibrating with home-made buffers that auditors reject.
- Demanding DRAP medical device registration for a laboratory pH meter, which does not apply and wrongly disqualifies bidders.