Specification builder · Tier A

Electronic pneumatic tourniquet system for limb surgery with automatic pressure regulation, single or dual channel, alarms, battery back-up, cuffs and optional limb occlusion pressure measurement

An electronic tourniquet inflates a cuff on the arm or leg and holds a set pressure so the surgeon can operate without bleeding. Orthopaedic, trauma, hand and paediatric theatres from THQ to tertiary level need one. The key tender decisions are one or two channels, accurate pressure control with loud alarms, a battery that survives load-shedding, a full range of cuffs at fixed prices, and whether built-in limb occlusion pressure measurement is worth the extra cost.

What this equipment will do

  • Patients: adults and children having upper-limb or lower-limb surgery, with cuffs from infant arm to large adult thigh, including contoured cuffs for conical limbs.
  • Bloodless surgical field: automatic inflation and regulation of cuff pressure over not more than 100 to not less than 450 mmHg, held within +/-6 mmHg or 2% of the set value and corrected for small leaks (IEC 60601-1).
  • Continuous display of set pressure, actual pressure and inflation time, with alarms for high or low pressure, leaks and elapsed inflation time (IEC 60601-1-8).
  • Keeps the cuff inflated through power cuts on an internal battery of not less than 4 hours.
  • Supports patient-specific pressure setting (limb occlusion or systolic pressure plus a margin) and short inflation times as advised by current tourniquet guidance (AORN 2025, BOAST 2021).
  • Two cuffs at the same time on two limbs (for example bilateral knee surgery) or on one limb, with separate pressure and timer for each. (with the optional module)
  • Built-in measurement of limb occlusion pressure with a recommended cuff pressure for each patient. (with the optional module)
  • IV regional anaesthesia (Bier block) with a double cuff and interlock against accidental deflation. (with the optional module)
  • Pre-use leak test of cuff, hose and connectors. (with the optional module)
  • Battery run time of not less than 8 hours. (with the optional module)
  • Export of a case and alarm log for audit. (with the optional module)

Not covered: It does not empty blood from the limb before inflation; that is done with an elastic exsanguination (Esmarch) bandage or an exsanguinator.; It is not an emergency or battlefield bleeding-control tourniquet; those are separate windlass or ratchet tourniquets.; It does not monitor the patient's blood pressure or oxygen levels; that needs a patient monitor.; It is not a blood-flow-restriction training device for physiotherapy unless the maker labels the model for that use.

9 international manufacturers checkedMarket checked 2026-1010 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 60601-1 General basic safety and essential performance of medical electrical equipment.
  • IEC 60601-1-2 Electromagnetic compatibility so the tourniquet keeps pressure near diathermy and other theatre equipment.
  • IEC 60601-1-8 Alarm systems: priorities, sounds and silencing of pressure and time alarms.
  • IEC 62304 Software life-cycle for the pressure-control and alarm software.
  • ISO 14971 Risk management for medical devices.
  • ISO 13485 Quality management system of the manufacturer.
  • IEC 62353 Recurrent and after-repair electrical safety testing of medical electrical equipment.
  • AORN Guideline for Pneumatic Tourniquet Safety US perioperative nursing guideline (approved April 2025) on cuff selection, patient-specific pressure, keeping inflation time short and documentation.
  • BOAST: The safe use of intraoperative tourniquets (BOA 2021) British Orthopaedic Association standard for safe tourniquet pressure (for example 50 mmHg above limb occlusion pressure in children), time and documentation.
  • DRAP Medical Devices Rules 2017 Registration or enlistment required to import and sell medical devices in Pakistan.

Common mistakes in tenders for this equipment

  • Copying one maker's brochure: a pressure range up to 700 mmHg, a named limb-occlusion method or a fixed screen size; these exclude equal products without clinical benefit, since safe surgery rarely needs more than 350-400 mmHg.
  • Buying the machine without a full cuff range: one adult arm and one thigh cuff cannot be used safely on children, obese patients or conical thighs.
  • Accepting units whose cuffs and connectors work only with the same maker's cuffs, then paying high prices for replacements; require standard connectors or supplied adapters and fixed cuff prices.
  • No acceptance check of displayed pressure against a calibrated pressure meter, so an inaccurate unit is used for years.
  • Ignoring the battery: a weak or missing battery deflates the cuff at every power cut; ask for automatic change-over, run time in hours and a replacement price.
  • Using the same high pressure for every patient instead of setting it from limb occlusion or systolic pressure; buy for training as well as hardware.
  • Accepting DRAP registration of a different model, or no service engineer in Pakistan for the offered model.