Specification builder · Tier B
Point-of-care (bedside) ultrasound for emergency, ICU, anaesthesia and obstetric triage: compact cart, laptop-style or handheld, with linear, convex and phased array coverage and needle guidance
A small, battery-powered ultrasound used by the treating doctor at the bedside to answer quick questions: bleeding in trauma, the heart and lungs in shock, fluid status, guiding nerve blocks and central lines, and checking the baby in labour room triage. Emergency departments, ICUs, operating theatres and labour rooms from THQ to tertiary level buy them, as compact touchscreen carts, laptop-style units or handheld probes that run on a phone or tablet. The key decisions are the format, the probe coverage, needle guidance tools, battery time, and for handhelds the subscription and display device.
What this equipment will do
- Patients: adults, children and pregnant women scanned at the bedside in emergency, ICU, theatres, labour rooms and wards, with acoustic output (MI/TI) shown on screen (IEC 60601-2-37, IEC 62359).
- Trauma bleeding (eFAST), abdomen, lung and pleura with the convex probe; veins, arteries, nerves and soft tissue with the linear probe; B-mode, M-mode, colour and power Doppler.
- Live ultrasound guidance of central lines, IV and arterial lines and nerve blocks with the linear probe.
- Needle guidance tools: on-screen guide lines, needle shaft highlighting, or needle tip tracking/navigation. (with the optional module)
- Focused heart scan (cardiac function, pericardial fluid) and inferior vena cava for fluid status with the phased array probe. (with the optional module)
- Blood velocity measurements by pulsed-wave Doppler: VTI, peak velocity and resistive index. (with the optional module)
- Heart valve and heart-muscle Doppler: continuous-wave Doppler up to 5 m/s and tissue Doppler. (with the optional module)
- Obstetric triage: fetal heart rate, presentation, gestational age and fetal weight, amniotic fluid. (with the optional module)
- Newborns and small children: small high-frequency linear or micro-convex probes for neonatal vascular access, lungs and abdomen. (with the optional module)
- Early pregnancy bleeding and gynaecology emergencies with the endocavity probe. (with the optional module)
- Heart scanning from inside the oesophagus (TEE) in theatre and ICU. (with the optional module)
- Automatic measurement helpers, as selected: ejection fraction, VTI or IVC collapse; lung B-line counting; bladder volume. (with the optional module)
Not covered: Full diagnostic radiology and echocardiography (detailed fetal anomaly scans, 3D/4D, breast, thyroid, comprehensive echo): need a general-purpose or cardiology ultrasound system.; Continuous fetal heart and contraction monitoring during labour (CTG): needs a fetal monitor.; Fractures, head injury bleeds and imaging through bone or bowel gas: need X-ray or CT.; Continuous vital-sign monitoring and alarms: need a patient monitor.
Loading the builder...
Standards this specification draws on
- IEC 60601-1 General electrical and mechanical safety of medical electrical equipment, including battery-powered and handheld units.
- IEC 60601-1-2 Electromagnetic compatibility, so images are not disturbed by diathermy, monitors, phones and wireless links, and the device does not disturb them.
- IEC 60601-2-37 Particular safety requirements for ultrasonic diagnostic and monitoring equipment, including acoustic output limits, probe surface temperature and on-screen MI/TI display.
- IEC 62359 Test methods for the thermal and mechanical indices shown to the operator.
- IEC 62133-2 Safety of the rechargeable lithium-ion batteries in the system and handheld probes.
- ISO 10993-1 Biological safety of transducer materials that touch skin, mucosa and sterile fields.
- IEC 62304 Software life-cycle control for the imaging, measurement and mobile application software.
- ISO 14971 Manufacturer's risk management for hazards such as probe overheating, wrong measurements, lost wireless link and data loss.
- ISO 13485 Certified quality management system of the manufacturer.
- DICOM (NEMA PS3) Standard for sending images and measurements to PACS and receiving worklists.
- IEC 62353 Electrical safety testing at installation, after repair and during preventive maintenance.
- DRAP Medical Devices Rules 2017 Pakistan's legal requirement for registration/enlistment and import of medical devices.
Common mistakes in tenders for this equipment
- Buying a handheld on the probe price alone: many need a paid yearly subscription and a compatible phone or tablet, so the device stops being useful when the first-year membership ends or the doctor changes phone.
- Letting scans live on personal phones or a foreign cloud account with no DICOM transfer, so images cannot be audited, reviewed or used for training and patient privacy is lost.
- Copying one maker's brochure: naming its needle tracking, automatic lung or cardiac tools or probe model numbers so only one bidder qualifies. Specify functions and frequency ranges with not less than values.
- Specifying battery life as standby time instead of continuous live scanning, then finding the unit dies halfway through a resuscitation.
- Ordering a single cart-based system for ER, ICU and OT together; it is never where it is needed. Several lower-cost laptop or handheld units often give more use.
- Leaving out sterile probe covers, needle guides, disinfection supplies and training, so nerve blocks and lines are done unsafely or not at all.
- Leaving probes, batteries and the supplied tablet out of the warranty and maintenance contract, although dropped probes and worn batteries are the commonest failures at the bedside.
- No hands-on training plan: POCUS without trained, credentialed users gives wrong decisions; budget training days and a phantom with the purchase.