Automatic digital polarimeter for optical rotation and specific rotation (589 nm, optional 546 / 365 nm, Peltier sample temperature control at 20 / 25 deg C)

An automatic polarimeter measures the optical rotation of solutions of sugars, antibiotics, amino acids and chiral drugs and calculates their specific rotation, an identity and purity test in many USP, BP and Ph. Eur. monographs. Pharma QC labs, drug testing labs, sugar and food labs and universities buy it. The real buying decisions are accuracy, built-in Peltier temperature control at 20 / 25 deg C, certified quartz plates for calibration and GMP data-integrity software.

What this equipment will do

  • Samples: solutions of APIs, excipients and raw materials that have a specific optical rotation limit in a monograph - sugars (sucrose, dextrose, lactose), antibiotics, amino acids, and chiral (single-enantiomer) drugs.
  • Sugars and syrups in sugar, food and excipient QC, read directly in the International Sugar Scale (deg Z) (OIML R 14). (with the optional module)
  • Optical rotation at 589 nm, the sodium D line (USP <781>, Ph. Eur. 2.2.7, BP Appendix V F, JP 2.49).
  • Specific rotation calculated from concentration, tube length and temperature, with pass / fail against monograph limits (USP <781>, Ph. Eur. 2.2.7).
  • Measurement at the pharmacopoeial temperature: 20 deg C (Ph. Eur. / BP) or 25 deg C (USP), with the sample temperature recorded.
  • Concentration of an optically active substance in solution from a known specific rotation.
  • Optical rotation at 546 nm (mercury green line) for methods that quote it. (with the optional module)
  • Optical rotation at several wavelengths from 365 to 589 nm (rotatory dispersion) for research. (with the optional module)
  • Instrument check of scale accuracy with certified quartz control plates and linearity with sucrose solutions (Ph. Eur. 2.2.7). (with the optional module)
  • Automatic check of tube filling for air bubbles or particles before measurement. (with the optional module)
  • Flow-through filling for series of similar samples without removing the tube. (with the optional module)

Not covered: Enantiomeric purity (amount of the wrong enantiomer) at low levels needs chiral HPLC on the HPLC system; optical rotation is not sensitive enough for that.; Refractive index and Brix of sugar solutions need the refractometer.; Circular dichroism spectra of proteins and chiral molecules need a circular dichroism spectrometer, not a polarimeter.; Melting point, assay and related substances of the same APIs need the melting point apparatus, HPLC and UV-Vis spectrophotometer.

8 international manufacturers checkedMarket checked 2026-1010 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 61010-1 Basic electrical and fire safety of laboratory equipment, including the polarimeter and its Peltier or water-bath temperature control.
  • IEC 61326-1 Electromagnetic compatibility of laboratory electrical equipment.
  • USP <781> USP optical rotation chapter: measurement at 589 nm and 25 deg C unless the monograph says otherwise, and calculation of specific rotation.
  • Ph. Eur. 2.2.7 European Pharmacopoeia optical rotation method: 589.3 nm at 20 deg C, scale checked with certified quartz plates and linearity with sucrose solutions.
  • BP Appendix V F British Pharmacopoeia optical rotation and specific optical rotation appendix, harmonised with Ph. Eur. 2.2.7 and cited in many Pakistani monographs.
  • JP 2.49 Japanese Pharmacopoeia general test 2.49 Optical Rotation Determination, used for products registered against JP monographs.
  • OIML R 14 OIML International Recommendation (current edition 1995) for polarimetric saccharimeters graduated in the ICUMSA International Sugar Scale (deg Z), relevant to sugar testing.
  • USP <1058> Framework for analytical instrument qualification (IQ/OQ/PQ).
  • 21 CFR Part 11 / EU GMP Annex 11 Electronic records, audit trail and electronic signature expectations for GMP labs inspected by DRAP.
  • ISO/IEC 17025 Competence of testing and calibration laboratories; requires traceable calibration of the polarimeter and quartz plates in accredited DTLs.

Common mistakes in tenders for this equipment

  • Copying one maker's brochure: an exact accuracy figure, number of wavelengths or named camera / chip recognition feature that only one maker offers, instead of performance requirements.
  • Asking for many wavelengths (365-880 nm) for routine pharma QC, where monographs use 589 nm; this narrows the field and adds cost with no benefit.
  • Leaving out certified quartz control plates with an accredited certificate, so the instrument cannot be verified at OQ or recalibrated later.
  • Not stating the temperature: Ph. Eur. / BP use 20 deg C and USP uses 25 deg C, and a polarimeter without controlled sample temperature gives results that fail audits.
  • Accepting "Part 11 compliant" claims with no audit trail or electronic signature demonstrated at OQ and no compliance matrix.
  • Forgetting spare tubes, windows and seals; a cracked window or leaking tube stops testing for weeks while parts are imported.
  • Asking for DRAP medical device registration, which does not apply to laboratory polarimeters and wrongly excludes valid bidders.