Hospital pressure swing adsorption (PSA) medical oxygen generation plant to ISO 10083, 250-1000 L/min, for pipeline supply with optional cylinder filling

A PSA oxygen plant compresses room air, dries and filters it and passes it through zeolite beds that hold back nitrogen, giving 93% oxygen straight into the hospital pipeline, with an optional booster that fills cylinders at about 150 bar. DHQ, tertiary and teaching hospitals in Pakistan buy them to stop depending on outside cylinder and liquid oxygen suppliers, a lesson from COVID-19. The key tender decisions are output at site conditions, back-up and reserve supply, feed-air quality, gas purity testing, cylinder filling safety, and the electricity, spares and operators needed to keep the plant running for 10 years.

What this equipment will do

  • Patients served: everyone on piped oxygen in ICU, HDU, emergency, theatres and wards, through the hospital medical gas pipeline (ISO 7396-1).
  • Makes oxygen continuously from room air: 250, 500 or 1000 L/min at 4-5 bar into the pipeline, rated at up to 45 deg C and 1500 m altitude (ISO 10083, WHO PSA specifications 2020).
  • Oxygen 93% (90-96%) meeting the pharmacopoeia Oxygen 93 per cent monograph, including limits for CO, CO2, water and oil (Ph. Eur. 2455). (with the optional module)
  • High-purity oxygen of 99% or more from a second polishing stage. (with the optional module)
  • Clean, dry feed air and oxygen-clean parts: compressor, dryer and multistage filters to ISO 8573-1, bacteria filter at the outlet (ISO 15001).
  • Continuous oxygen purity, pressure and flow display with low-purity alarm at 90% and automatic block so off-specification gas cannot enter the pipeline (ISO 10083).
  • Automatic switch-over to the reserve supply with alarm when the plant stops or pressure or purity falls, and automatic restart after power failure (ISO 7396-1).
  • Two complete independent trains, each able to carry the full load, with automatic standby start and maintenance without stopping supply. (with the optional module)
  • Filling medical oxygen cylinders at about 150 bar, about 35 or 85 cylinders a day, with pipeline supply always given priority (ISO 15001, ISO 9809-1). (with the optional module)
  • Automatic reserve cylinder manifold with duty and standby banks and alarms. (with the optional module)
  • Continuous monitoring of carbon monoxide, carbon dioxide and dew point in the product oxygen. (with the optional module)
  • Plant alarms repeated to a 24-hour staffed room, the building management system or mobile phones. (with the optional module)

Not covered: Liquid oxygen, or guaranteed 99.5% pharmacopoeial Oxygen for every use: needs a liquid oxygen (VIE) tank supplied by a gas company.; Medical breathing air for ventilators and surgical air for tools: needs a medical air compressor plant.; Suction (medical vacuum) for wards and theatres: needs a medical vacuum plant.; The pipeline network, ward outlets, area valves and area alarms: are part of the medical gas pipeline system works (ISO 7396-1); the plant only connects to it.

8 international manufacturers checkedMarket checked 2026-1011 standards referenced

Loading the builder...

Standards this specification draws on

  • ISO 10083 Particular standard for oxygen concentrator (PSA) supply systems feeding medical gas pipelines: minimum oxygen concentration, sources of supply, analysers, alarms, testing and commissioning.
  • IEC 60601-1 Basic safety and essential performance of the medical electrical control, monitoring and alarm equipment of the plant.
  • ISO 7396-1 Medical gas pipeline systems: primary, secondary and reserve supply, alarms, pipeline connection and commissioning tests that the plant must fit into.
  • ISO 8573-1 Purity classes for compressed air, used to set the particle, water and oil quality of feed air entering the PSA generator.
  • Ph. Eur. monograph 2455 Oxygen (93 per cent) Pharmacopoeial quality of oxygen 93% (also in the British Pharmacopoeia; USP has Oxygen 93 Percent): oxygen content and limits for CO, CO2, water and oil.
  • ISO 15001 Compatibility with oxygen and cleanliness of components in oxygen service, critical for the booster, filling ramp and receivers.
  • ISO 9809-1 Design, construction and testing of refillable seamless steel gas cylinders used with the filling station and reserve manifold.
  • ISO 13485 Quality management system of the legal manufacturer of the medical oxygen generator.
  • ISO 14971 Manufacturer's risk management for hazards such as low purity, contaminated gas, fire and loss of supply.
  • WHO Technical specifications for PSA oxygen plants (2020) WHO minimum specifications for PSA oxygen plants supplying medical-grade oxygen in low- and middle-income countries.
  • DRAP Medical Devices Rules 2017 Legal requirement for registration or enlistment of medical devices imported or sold in Pakistan.

Common mistakes in tenders for this equipment

  • Copying one maker's model figures, such as an exact output in Nm3/h, receiver volume or controller feature, so that only one plant fits. Use 'not less than' output at stated site conditions and accept any design that meets ISO 10083.
  • Sizing the plant on bed numbers only. ICU ventilators and high-flow nasal oxygen use 15-60 L/min per patient, and a cylinder booster takes its flow from the same plant.
  • Accepting output and purity measured at the factory at 20 deg C. Demand derated figures for 45 deg C and the site altitude, and a 72-hour performance run at site before acceptance.
  • Buying a single-train plant with no automatic reserve manifold or generator connection, so oxygen stops during load-shedding or the first compressor fault.
  • Accepting an industrial oxygen generator with an industrial CE declaration, no ISO 13485 and no DRAP registration, often from the same factories that make fish-farm and welding units.
  • Filling cylinders with a lubricated booster, non-cleaned pigtails or untested cylinders, which risks fire and cylinder bursts.
  • No budget for electricity, filters, analyser sensors, trained operators or AMC, so plants installed during COVID-19 stood idle after the first fault.
  • Placing the air intake near generator exhausts, kitchens or roads and never testing product gas for carbon monoxide, carbon dioxide and moisture.