Specification builder · Tier C

Large horizontal pre-vacuum steam sterilizer for hospital CSSD and operating theatres (chamber 2 to 12+ sterilisation units, EN 285)

A large, floor-standing autoclave that uses pressurised saturated steam, with pulsed vacuum air removal, to sterilise wrapped surgical instruments, containers, textiles and other heat-resistant items. It is the main machine of a hospital Central Sterile Services Department (CSSD) and of operating theatre sterile cores. Capacity is stated in sterilisation units (STU, one 300 x 300 x 600 mm basket), from about 2 STU for a theatre suite up to 12 STU or more for a large teaching hospital.

What this equipment will do

  • Loads: wrapped surgical instrument sets, rigid containers and textile packs (gowns, drapes), 2 to 12+ sterilisation units per cycle, on pre-vacuum 134 deg C cycles (EN 285, ISO 17665).
  • Loads: heat-sensitive reusable rubber and plastic items on a gentler 121 deg C cycle (EN 285).
  • Removes air from packs and hollow instruments by repeated vacuum pulses and dries the load by vacuum at the end, so packs come out dry (EN 285 drying test).
  • Daily machine checks: Bowie-Dick steam penetration test and automatic vacuum leak test programmes (EN 285, EN ISO 11140-4).
  • Records every cycle (programme, temperature, pressure, operator, pass/fail) for load release and audit, with temperature and pressure kept within the EN 285 sterilisation band (ISO 17665).
  • Detects air or non-condensable gas left in the chamber and flags the cycle (EN 285 air detector), or monitors steam penetration in every cycle. (with the optional module)
  • Extended 134 deg C 18-minute cycle for prion-risk instruments, rapid cycle for unwrapped urgent instruments, and extra-drying cycle for rigid containers. (with the optional module)
  • Liquids in open or vented containers, with slow cooling to stop boiling over (ISO 17665). (with the optional module)
  • Pass-through barrier layout: dirty side loads, sterile side unloads, and the sterile door opens only after a successful cycle. (with the optional module)
  • Makes its own steam from electricity where the hospital has no boiler steam, or keeps working on electric steam when the boiler fails. (with the optional module)
  • Tracks each instrument set by barcode from washer and sterilizer cycle to expiry date and the patient record. (with the optional module)

Not covered: Heat-sensitive devices that cannot take steam (flexible endoscopes, many electronic and plastic devices) need a low-temperature sterilizer (hydrogen peroxide or ethylene oxide) or an endoscope washer-disinfector.; Cleaning and disinfecting used instruments before packing needs a washer-disinfector (ISO 15883) and ultrasonic cleaner; the sterilizer does not clean.; Small loads in clinics, dental units and wards (under one sterilisation unit) need a bench-top Class B sterilizer (EN 13060).; Decontaminating laboratory waste and sterilising culture media need a laboratory autoclave.

8 international manufacturers checkedMarket checked 2026-1010 standards referenced

Loading the builder...

Standards this specification draws on

  • EN 285:2015+A1:2021 The European performance and test standard for large steam sterilizers (1 STU and above); defines STU, air detector, leak test, thermometric and drying tests.
  • ISO 17665:2024 Requirements for development, validation (IQ/OQ/PQ) and routine control of moist heat sterilisation processes for health care products.
  • EN 13060 Standard for small steam sterilizers (below 1 STU); bidders must not offer EN 13060 bench-top units for a CSSD large-sterilizer item.
  • PED 2014/68/EU / ASME BPVC Section VIII Div. 1 Pressure vessel design, manufacture and certification of the chamber, jacket and steam generator.
  • IEC 61010-1 and IEC 61010-2-040 Electrical and mechanical safety of sterilizers and washer-disinfectors.
  • IEC 61326-1 EMC requirements for laboratory and process control equipment, applies to sterilizer controls.
  • EN ISO 11140-1 / -4 Chemical indicators and Bowie-Dick type test packs used for daily testing and load monitoring.
  • EN ISO 11138-1 / -3 Biological indicators for steam sterilisation used in validation and routine monitoring.
  • EN ISO 13485:2016 Quality management system for medical device manufacturers.
  • ANSI/AAMI ST79 / UK HTM 01-01 Widely used guidance for steam sterilisation practice, testing frequency and CSSD design.

Common mistakes in tenders for this equipment

  • Writing exact chamber dimensions or cabinet width (e.g. one maker's 660 x 660 x 1000 mm chamber) instead of a minimum STU and usable volume, which excludes equivalent machines.
  • Naming proprietary features or sensors (branded air-removal sensors, branded water-saving or software names) instead of describing the function and the measurable performance.
  • Mandating one air-removal technology (only liquid-ring pump, or only ejector) when the EN 285 tests are what prove performance.
  • Requiring compatibility with an existing brand's loading carts or traceability software without allowing an open data interface, which locks future purchases.
  • Accepting an EN 13060 bench-top or a 'meets EN 285 test methods' self-declaration in place of EN 285 type-test evidence for a CSSD machine.
  • Leaving RO water treatment, loading carts, barrier-wall flange, voltage stabiliser and validation out of scope, so the lowest bidder later adds them as variations.
  • Comparing only purchase price and ignoring water, electricity, and the after-warranty CMC price, which over 10 years can exceed the machine cost.
  • Not asking the bidder to state full-load cycle time including drying, so a cheap machine with half the daily throughput wins.
Medicine brands A–Z:ABCDEFGHIJKLMNOPQRSTUVWXYZ0-9