Sterility testing isolator with integrated H2O2 bio-decontamination, Grade A airflow and integrated sterility test pump
A sterility testing isolator is a sealed, glove-operated Grade A work chamber in which injectables, eye drops and sterile medical devices are tested for microbial contamination by the USP <71> / Ph. Eur. 2.6.1 sterility test. Its built-in hydrogen peroxide cycle sterilises the inside and the outside of loaded items, so the test is far less likely to give false positives than an open laminar-flow hood in a Grade B cleanroom. It is bought by drug testing laboratories, pharmaceutical QC labs releasing sterile products and medical-device makers.
What this equipment will do
- Products: injectables, eye drops, biologicals and other sterile medicines (USP <71>, Ph. Eur. 2.6.1).
- Products: sterile medical devices such as syringes, IV sets, catheters, sutures and gloves (USP <71>, ISO 11737-2).
- Sterility test by closed-system membrane filtration with a sterility test pump inside the Grade A chamber (USP <71>, Ph. Eur. 2.6.1).
- Two independently controlled pumps, so two tests can be filtered at the same time. (with the optional module)
- Hydrogen peroxide bio-decontamination of the chamber, transfer hatch and outside of loaded items to 6-log reduction of biological indicators, with aeration to 1 ppm or less (ISO 13408-6).
- Grade A / ISO 5 air inside the chamber through H14 HEPA filters (EU GMP Annex 1, ISO 14644-1).
- Positive chamber pressure with alarm, and chamber leak test by pressure decay (ISO 14644-7, ISO 10648-2).
- Glove integrity testing from outside without breaking the sterile barrier (EU GMP Annex 1).
- Loading of the next samples while testing continues, through a separately decontaminated transfer hatch. (with the optional module)
- Aseptic transfer of pre-sterilised containers or rinse fluids without opening the barrier, through a rapid transfer port or sterile liquid port. (with the optional module)
- Particle counting of the chamber air to show Grade A during testing (ISO 14644-1). (with the optional module)
- Electronic cycle records with audit trail and electronic signatures (21 CFR Part 11, EU GMP Annex 11). (with the optional module)
Not covered: It does not incubate the media or read the result; the canisters need incubators at 20-25 deg C and 30-35 deg C for 14 days (USP <71>).; It does not test bacterial endotoxins or pyrogens; that needs the endotoxin (LAL) test system.; It does not identify the organism found in a positive test; that needs a microbial identification system.; It is not a production isolator for aseptic filling, nor a negative-pressure containment isolator for handling cytotoxic or hazardous drugs.
Loading the builder...
Standards this specification draws on
- USP <71> Sterility Tests chapter (harmonised with Ph. Eur. 2.6.1 and JP 4.06) that the isolator exists to perform; it allows testing in an isolator or in Grade A within a Grade B cleanroom.
- Ph. Eur. 2.6.1 European Pharmacopoeia sterility test, used through the BP in Pakistan; requires aseptic conditions and method suitability testing.
- ISO 13408-6:2021 Aseptic processing of health care products - Part 6: isolator systems; sets design, decontamination, leak and qualification requirements.
- EU GMP Annex 1 (2022) EU / PIC/S GMP Annex 1 (PIC/S PE 009 Annex 1, 2022) on sterile manufacture: isolator background grade, glove integrity testing, sporicidal decontamination and Grade A limits.
- ISO 14644-7:2004 Cleanrooms - separative devices (isolators, glove boxes); leak testing and access device requirements.
- ISO 14644-1:2015 Classification of air cleanliness by particle concentration; used to prove ISO 5 / Grade A inside the isolator.
- USP <1208> Sterility Testing - Validation of Isolator Systems: guidance on isolator qualification, decontamination validation and glove/leak integrity.
- ISO 11737-2:2019 Tests of sterility performed in the validation and maintenance of medical-device sterilization processes; relevant for device makers' sterility testing.
- 21 CFR Part 11 US FDA rule on electronic records and signatures; applies to the isolator's cycle, alarm and audit-trail records.
- IEC 61010-1 Electrical safety standard for laboratory equipment, including the isolator and its peroxide generator.
- IEC 61326-1:2020 EMC requirements for electrical equipment for measurement, control and laboratory use.
- ISO/IEC 17025:2017 Accreditation standard for testing labs; requires qualified equipment and traceable calibration.
Common mistakes in tenders for this equipment
- Buying a Grade A laminar-flow hood for a Grade B cleanroom to save money: the cleanroom suite, airlocks, HVAC, gowning, daily environmental monitoring and higher false-positive rate usually cost more over 5 years than an isolator in a Grade D room, and EU GMP Annex 1 favours isolators.
- Accepting a '6-log decontamination' claim from a brochure instead of requiring site cycle development with biological indicators at worst-case locations and three consecutive successful runs.
- Naming one maker's peroxide technology (vapour, ionised mist or dry fog), chamber size or glove count; specify the performance (6-log BI reduction, aeration to not more than 1 ppm, Grade A, internal dimensions 'not less than') instead.
- Forgetting that the sterility test pump decides which filtration canisters you can buy for the next 10 years; ask bidders to state canister compatibility and price consumables for 5 years.
- Leaving out BIs, sterilant, spare gloves, HEPA filters and canisters, so the isolator cannot be qualified or stops after the first glove tear.
- Ignoring residual hydrogen peroxide: poorly aerated loads can kill organisms in the test and give false 'sterile' results; require aeration limits and growth-promotion / method suitability checks with the decontaminated load.
- Not checking the room before ordering: the isolator needs a Grade D or controlled background, exhaust for peroxide, enough ceiling height and floor loading, and sometimes 3-phase power and compressed air.
- Asking for DRAP medical-device registration: a sterility testing isolator is laboratory equipment, so this demand only confuses bidders.