Specification builder · Tier A
Closed-system sterility testing pump with single-use canisters and multi-place stainless-steel membrane filtration manifold with vacuum source
A closed-system sterility testing pump pushes an injectable, rinse fluids and culture media through sealed single-use canisters with 0.45 um membranes, so the sterility test of USP <71> / Ph. Eur. 2.6.1 is done without opening the sample to room air. A stainless-steel manifold with 3 or 6 funnels on a vacuum pump is used for membrane-filtration microbial counts of non-sterile products, device bioburden and water. Drug and device testing laboratories, pharmaceutical QC labs, device manufacturers and universities buy them; most of the lifetime cost is in canisters, funnels and membranes.
What this equipment will do
- Sterility samples: filterable injectables and other aqueous liquids in vials, ampoules, bottles and bags (USP <71>, Ph. Eur. 2.6.1). (with the optional module)
- Sterility samples: antibiotics and powders to be dissolved, with rinsing to remove the product's antimicrobial activity. (with the optional module)
- Sterility samples: medical devices and prefilled syringes, including flushing of device lumens. (with the optional module)
- Closed-system membrane filtration sterility test: equal split to two canisters with 0.45 um membranes, rinsing and media addition without opening the sample to room air (USP <71>, Ph. Eur. 2.6.1). (with the optional module)
- Line pressure monitoring with alarm and automatic slowing or stopping of the pump to protect the membranes, and stored test methods. (with the optional module)
- Use inside a sterility testing isolator (peroxide-resistant, operated through the gloves). (with the optional module)
- Open-method sterility test on a stainless-steel manifold with a sterile rinse line (USP <71>). (with the optional module)
- Microbial counts of non-sterile medicines and raw materials by membrane filtration (USP <61>, Ph. Eur. 2.6.12) and tests for specified micro-organisms (USP <62>). (with the optional module)
- Bioburden of medical devices by filtering device eluates (ISO 11737-1). (with the optional module)
- Water testing: total counts in purified and drinking water and E. coli / coliform counts by membrane filtration (ISO 8199, ISO 9308-1). (with the optional module)
- Electronic run records with audit trail and electronic signatures (21 CFR Part 11, EU GMP Annex 11). (with the optional module)
Not covered: It does not provide the aseptic environment itself; the sterility test must be done in a sterility testing isolator or a Grade A cabinet inside a Grade B cleanroom.; It does not incubate or read the tests; canisters and membranes need incubators (20-25 deg C and 30-35 deg C).; It does not test bacterial endotoxins; that needs the endotoxin (LAL) test system.; It does not count particles in injections; that needs a light-obscuration particle counter (USP <788>).
Loading the builder...
Standards this specification draws on
- IEC 61010-1 Basic electrical safety standard for laboratory equipment; applies to the sterility pump and the vacuum pump used in wet cleanroom areas.
- IEC 61326-1 EMC standard for laboratory equipment, so the pump electronics work reliably next to other cleanroom equipment.
- USP <71> Sterility Tests chapter; defines the membrane filtration method, 0.45 um membranes, rinsing and method suitability that the pump and canisters must support.
- Ph. Eur. 2.6.1 European (and BP) sterility test chapter, harmonised with USP <71>; used for BP monograph products in Pakistan.
- USP <61> Microbial enumeration tests for non-sterile products; membrane filtration on the manifold is one of its counting methods.
- USP <62> Tests for specified micro-organisms in non-sterile products, often run with membrane filtration on the same manifold.
- Ph. Eur. 2.6.12 European (and BP) microbial enumeration chapter, harmonised with USP <61>.
- ISO 11737-1:2018 Bioburden determination on medical devices, where membrane filtration of eluates is a common counting method (with Amd 1:2021).
- ISO 8199:2018 General requirements and guidance for microbial counts in water by culture, including membrane filtration.
- ISO 9308-1:2014 Enumeration of E. coli and coliform bacteria in water by membrane filtration (with Amd 1:2016).
- 21 CFR Part 11 US FDA rule on electronic records and signatures; applies when pump run records are kept electronically for release decisions.
- ISO/IEC 17025:2017 Competence standard for testing and calibration laboratories; drives the need for traceable calibration of speed, time, pressure and vacuum.
Common mistakes in tenders for this equipment
- Specifying the pump by a brand's canister design (needle shape, port layout or tubing set) so that only one maker's canisters fit, which locks the lab into one consumable supplier for the life of the pump.
- Buying the pump without any canisters, rinse fluids or media, so it sits unused until a separate consumables tender is finished.
- Evaluating only the pump price and ignoring the price of canisters, which over five years usually costs several times more than the pump.
- Not stating the product types to be tested (antibiotics, powders, oily products, devices, prefilled syringes), so the supplied canisters cannot handle part of the workload.
- Asking for an isolator-ready pump without checking the isolator wall cut-out, glove reach and decontamination cycle, so the pump cannot be installed or fails after repeated peroxide cycles.
- Supplying a manifold with plastic valves or seals that crack after a few autoclave cycles, or a manifold too tall to work under the laminar-airflow cabinet.
- Using an oil-sealed or undersized vacuum pump without a waste trap and filter, so liquid and microbes are sucked into the pump and vacuum is too weak for six places.
- Accepting no method training or method suitability support, leading to false-positive sterility failures and repeated tests.