Specification builder · Tier C
Liquid-borne light-obscuration particle counter for sub-visible particles in injections, infusions, ophthalmic solutions and medical-device rinse fluids (USP <788>/<787>/<789>, Ph. Eur. 2.9.19, JP 6.07)
A sub-visible particle counter counts and sizes particles too small to see (mainly 10 um and 25 um and larger) in injections and infusions by passing an exact volume of liquid through a light beam, as required by the pharmacopoeias. It is bought by drug and device testing laboratories, manufacturers of infusions, ampoules, vials, eye drops and IV sets, and biologics labs. The membrane microscope method is the backup for products that cannot be tested this way.
What this equipment will do
- Samples: large-volume infusions (bags and bottles of 100 mL or more) and small-volume injections in ampoules, vials and prefilled syringes, including powders after reconstitution (USP <788>, Ph. Eur. 2.9.19, JP 6.07).
- Samples: rinse fluid from infusion sets, syringes and catheters, supporting the particulate contamination and cleanliness requirements of ISO 8536-4 and ISO 7886-1. (with the optional module)
- Samples: therapeutic protein injections in small volumes of 1 mL or less (USP <787>). (with the optional module)
- Samples: eye drops and ophthalmic solutions, including the 50 um size class (USP <789>). (with the optional module)
- Light-obscuration particle count: particles of 10 um and 25 um and larger per mL (infusions) or per container (small-volume injections), with pass/fail against the pharmacopoeial limits (USP <788>, Ph. Eur. 2.9.19, JP 6.07).
- Particle size distribution from about 2 um to 100 um in user-set size classes.
- Large particles and fibres up to 400 um in device rinse fluid. (with the optional module)
- Viscous products up to about 40 cP without dilution. (with the optional module)
- Particle images to tell protein aggregates, silicone oil droplets, air bubbles and fibres apart. (with the optional module)
- Checks of sizing, resolution and counting accuracy with certified standards (USP <788>, ISO 21501-3). (with the optional module)
- Filtering samples onto membranes for the backup membrane microscope count (USP <788> method 2); the microscope count itself needs a separate microscope. (with the optional module)
- Electronic records with audit trail and electronic signatures (21 CFR Part 11, EU GMP Annex 11). (with the optional module)
Not covered: It does not inspect for visible particles in filled containers; that needs a visual inspection booth or automated inspection machine (USP <790>, Ph. Eur. 2.9.20).; It does not count particles in cleanroom air; that needs an airborne particle counter (ISO 14644-1).; It does not measure the particle size of powders, suspensions or emulsions; that needs a laser diffraction particle size analyser.; It does not perform the membrane microscope count by itself; that needs a binocular microscope with an ocular micrometer, which is not included.
Loading the builder...
Standards this specification draws on
- USP <788> Particulate Matter in Injections: defines the light-obscuration test, its calibration (sizing, resolution, counting and volume accuracy) and the limits for particles of 10 um and 25 um and larger.
- Ph. Eur. 2.9.19 European (and BP-harmonised) sub-visible particle test used for most injectables registered in Pakistan.
- JP 6.07 Japanese Pharmacopoeia insoluble particulate matter test for injections, harmonised with USP <788> and Ph. Eur. 2.9.19.
- USP <787> Subvisible particulate matter in therapeutic protein injections; allows reduced test volumes.
- USP <789> Particulate matter in ophthalmic solutions, including a 50 um size threshold.
- ISO 21501-3:2019 Calibration and verification method for light-extinction liquid-borne particle counters.
- ISO 8536-4:2019 Infusion sets for single use, gravity feed; includes the particulate contamination requirement for infusion sets.
- ISO 7886-1:2017 Sterile hypodermic syringes for single use; includes the cleanliness requirement for syringes.
- 21 CFR Part 11 US FDA rule on electronic records and signatures that the particle counting software is checked against.
- ISO/IEC 17025:2017 Accreditation standard for testing labs; requires calibrated equipment and controlled data.
- IEC 61010-1 Electrical safety standard for laboratory equipment (counter, clean bench, ultrasonic bath).
- IEC 61326-1 EMC requirements for laboratory electrical equipment.
Common mistakes in tenders for this equipment
- Copying one maker's sensor range, channel count or flow rate (for example an exact lower size limit or 1,000+ channels) so that only one bidder can comply; specify the pharmacopoeial calibration requirements instead.
- Accepting a 'meets USP <788>' claim without an installation check of sizing, resolution, counting accuracy and sample volume accuracy with certified standards.
- Buying the counter without certified size and count standards, particle-free water filtration, clean glassware and a laminar-flow bench, so blanks fail and results cannot be trusted.
- Forgetting that the sensor must be calibrated every year: if it has to be shipped abroad, testing stops for weeks unless on-site calibration or a loan unit is written into the contract.
- Expecting light obscuration to test every product: emulsions, colloids and viscous products need the membrane microscope method, so the lab still needs a membrane set and microscope.
- Accepting a '21 CFR Part 11 compliant' claim without a live demonstration of audit trail, user roles, electronic signatures and protection against deletion of raw count data.
- Asking for DRAP medical-device registration for the counter: it is laboratory equipment, and the demand only confuses bidders.