Surgical smoke (plume) evacuation unit with ULPA and activated carbon filtration, automatic activation with electrosurgery, and capture accessories for open and laparoscopic surgery

A smoke evacuator removes and filters the smoke made by diathermy and lasers during surgery, protecting theatre staff from breathing fine particles, microbes and chemicals. It is bought for operating theatres from DHQ to tertiary level and for laser, dermatology and LEEP rooms. The key tender decisions are ULPA plus carbon filtration, automatic start with the theatre's existing generators, open versus laparoscopic suction, and above all the filter life and price, which decide the running cost.

What this equipment will do

  • Smoke from electrosurgery (diathermy), surgical lasers and ultrasonic instruments, captured at the source (ISO 16571).
  • Filtration: pre-filter, ULPA filter capturing not less than 99.999% of fine particles in the 0.1-0.2 um range, and activated carbon for odour and gases (EN 1822-1 method for filter efficiency).
  • Capture in open surgery through smoke-evacuation pencils or a wand held close to the tissue, with suction adjustable in steps. (with the optional module)
  • Low-flow laparoscopic smoke removal from a trocar or port without deflating the abdomen. (with the optional module)
  • Automatic start and stop when the diathermy pencil or laser is used, with generators of any make. (with the optional module)
  • Automatic filter-life tracking and change-filter warning. (with the optional module)
  • Highest filter class: ULPA 99.9995% (U15). (with the optional module)
  • High free-air flow (not less than 700 L/min) for heavy laser or diathermy plume. (with the optional module)
  • Quiet running at not more than 58 dB(A) at maximum suction. (with the optional module)
  • Automatic run-on of up to 10 seconds to clear residual smoke. (with the optional module)

Not covered: It does not remove anaesthetic gases; that is done by the anaesthetic gas scavenging system of the anaesthesia machine.; It is not a fluid or blood suction pump; theatre suction or a fluid waste management system does that.; It does not replace theatre ventilation (air changes and HEPA supply of the OT) or masks; it only removes smoke at the source.; It does not cut or coagulate tissue; that needs an electrosurgical unit, laser or ultrasonic device.

9 international manufacturers checkedMarket checked 2026-109 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 60601-1 General basic safety and essential performance of medical electrical equipment.
  • IEC 60601-1-2 Electromagnetic compatibility, important because the unit sits beside, and senses, the electrosurgical generator.
  • ISO 16571 ISO 16571:2024 (with Amd 1:2026), the international standard for systems that evacuate plume generated by medical devices: design, testing, filtration, installation and maintenance.
  • EN 1822-1 Classification and testing of EPA, HEPA and ULPA filters, used to state and check the ULPA filter efficiency.
  • ISO 14971 Risk management for medical devices.
  • ISO 13485 Quality management system of the manufacturer.
  • IEC 62353 Recurrent and after-repair electrical safety testing of medical electrical equipment.
  • AORN Guideline for Surgical Smoke Safety Perioperative guideline on evacuating surgical smoke at the source, filter choice, suction and capture-device position.
  • DRAP Medical Devices Rules 2017 Registration or enlistment required to import and sell medical devices in Pakistan.

Common mistakes in tenders for this equipment

  • Copying one maker's figures (a fixed flow in L/min, a filter-life hour count or a trademarked filter name) instead of asking for ULPA efficiency, carbon stage, suction steps and stated flow; this excludes equal products.
  • Accepting automatic activation that works only with the bidder's own generator, so the evacuator does not start with the theatre's existing diathermy machines.
  • Buying the unit without a year's filters, pencils and tubing, or with filters, pencils and tubing that only one source can supply.
  • Judging bids on unit price only: filter life varies from about 5 to 40 hours between makers, so filter cost per hour decides the 5-year cost.
  • Using wall suction or a general vacuum instead of an evacuator: it lacks ULPA and carbon filtration and blocks the hospital vacuum line.
  • No training on holding the capture device within about 5 cm of the smoke source and on handling used filters as contaminated waste, so the unit runs but captures little.
  • Accepting DRAP registration of a different model, or no local service engineer and filter stock in Pakistan.