Specification builder · Tier A
Temperature (and temperature/humidity) data loggers and mapping kit for storage areas, laboratory equipment and sterilisers, with calibration certificates and validation software
A mapping kit is a set of calibrated, battery-powered data loggers placed at many points in a warehouse, cold room, refrigerator, freezer, stability chamber, oven or steriliser to record temperature (and humidity) and find hot and cold spots. Pharma manufacturers, distributors, drug testing labs and hospital pharmacies need it to qualify storage and equipment under GMP/GDP and WHO guidance. The real buying decisions are temperature range, accuracy class, accredited calibration of every logger, number of loggers and whether the software has GMP data-integrity controls.
What this equipment will do
- Rooms and stores: warehouses, cold rooms and medicine stores mapped on a grid at several heights (WHO TRS 961 Annex 9, Supplement 8).
- Equipment from -30 to +70 deg C: refrigerators, stability chambers, incubators and lab rooms.
- Ultra-low and -20 deg C freezers down to -80 deg C. (with the optional module)
- Drying ovens and other hot equipment up to +140 deg C. (with the optional module)
- Steam sterilisers and autoclaves: chamber and load temperature with chamber pressure (ISO 17665, EN 285). (with the optional module)
- Dry-heat sterilisers and depyrogenation ovens or tunnels up to about 300 deg C. (with the optional module)
- Temperature recorded at every position at a set interval, from 1 minute (storage) down to 1 second (door-opening and sterilisation studies). (with the optional module)
- Relative humidity at selected positions in stability chambers, stores and lab rooms. (with the optional module)
- Mapping analysis: minimum, maximum and mean per position, warmest and coldest positions, and time outside limits. (with the optional module)
- F0 lethality of each position over a steam sterilisation cycle. (with the optional module)
- Live view of a running study and alarms when limits are exceeded. (with the optional module)
- Spot checks of chamber, refrigerator and oven display readings with a calibrated reference thermometer. (with the optional module)
- Daily temperature and humidity records of laboratory rooms with displayed thermo-hygrometers. (with the optional module)
- In-house check of the loggers against a reference before and after each study. (with the optional module)
Not covered: Permanent 24/7 monitoring of many refrigerators and rooms with SMS alarms to staff is a separate system: see the remote temperature monitoring system.; Accredited calibration of the loggers themselves needs a calibration laboratory with reference baths and standards; the kit only carries certificates and, optionally, an in-house check.; Proof that a sterilisation cycle actually kills microorganisms needs biological indicators (and an indicator incubator); air-removal checks need Bowie-Dick test packs.; Clean-room airflow, particle counts and room pressure differentials need an anemometer, particle counter and differential pressure meter.
Loading the builder...
Standards this specification draws on
- IEC 61010-1 Basic electrical and fire safety of laboratory and measurement equipment; applies to mains-powered readers, base stations, reference thermometers and calibration baths of the kit.
- WHO TRS 961 Annex 9, Supplement 8 WHO technical supplement 'Temperature mapping of storage areas': logger accuracy (+/-0.5 deg C), 3-point calibration, 1-15 minute interval, logger layout, study duration, seasonal mapping and report contents.
- IEC 61326-1 Electromagnetic compatibility of electrical measurement and laboratory equipment, including radio-linked loggers and base stations.
- EN 12830 European standard for temperature recorders used in the storage and distribution of chilled and frozen goods (tests, accuracy class and suitability).
- ISO/IEC 17025 Competence of calibration laboratories; an accredited certificate gives each logger's error and uncertainty with SI traceability.
- 21 CFR Part 11 / EU GMP Annex 11 Electronic records, audit trail and electronic signature expectations for mapping and qualification data in GMP facilities inspected by DRAP.
- EU GMP Annex 15 Qualification and validation of facilities, equipment and processes, including temperature-controlled equipment and sterilisers.
- ISO 17665 Moist heat sterilisation of health care products: development, validation and routine control, including temperature and pressure measurement in the chamber and load.
- EN 285 Large steam sterilisers: thermometric and pressure test methods used when validating hospital and industrial autoclaves.
- IEC 60751 Industrial platinum resistance thermometers and their tolerance classes, used to define sensor quality of validation loggers and reference probes.
Common mistakes in tenders for this equipment
- Copying one maker's brochure: exact logger dimensions, named radio systems or a specific software name instead of range, accuracy, memory and data-integrity requirements.
- Accepting a 'calibration certificate' that is only a factory test at one point, or 'traceable' wording without an ISO/IEC 17025 accredited certificate per logger at 3 points across the range of use.
- Buying storage loggers rated only to +60 or +70 deg C and then using them in ovens or autoclaves, where they fail or leak.
- Forgetting stands, holders, spare batteries and a reference thermometer, so loggers end up on the floor or flat in the middle of a 7-day study.
- Asking for real-time wireless and remote alarms when the need is periodic mapping; it narrows the field and adds radio approval and IT costs. Continuous monitoring belongs in a separate monitoring system tender.
- Accepting 'Part 11 compliant' claims without user log-ins, audit trail and electronic signatures demonstrated at OQ.
- Mixing logger types from different makers in one study, against WHO advice, because the kit was too small and extra loggers were bought later.
- Asking for DRAP medical device registration, which does not apply to measuring instruments and wrongly excludes valid bidders.