External temporary cardiac pacemaker (single- or dual-chamber external pulse generator) for transvenous and epicardial pacing in CCU, ICU, cath lab and cardiac surgery
A temporary cardiac pacemaker is a small battery-powered box that sends electrical pulses to the heart through a wire passed through a vein (transvenous) or through wires stitched to the heart surface during cardiac surgery (epicardial). CCUs, ICUs, emergency departments, cath labs and cardiac surgery units buy them to keep patients with dangerously slow rhythms or heart block alive until the rhythm recovers or a permanent pacemaker is implanted. In Pakistan the key tender decisions are single- versus dual-chamber units, patient cables and lead consumables, a spare unit for pacing-dependent patients, and fast local service.
What this equipment will do
- Patients: adults and children with dangerously slow heart rhythms or heart block (for example after a heart attack, drug toxicity or before a permanent pacemaker), paced through a temporary transvenous catheter (IEC 60601-2-31).
- Patients after open-heart surgery, paced through temporary epicardial heart wires placed by the surgeon (IEC 60601-2-31).
- Single-chamber pacing of the atrium or ventricle, on demand (VVI/AAI, pacing only when the heart's own beat is missing) or at a fixed rate (VOO/AOO), rate at least 30 to 200 per minute.
- Dual-chamber pacing that keeps atria and ventricles beating in sequence (DDD and related modes), with separate atrial and ventricular output and sensitivity and adjustable AV delay. (with the optional module)
- Biventricular temporary pacing of both ventricles after cardiac surgery, through a third output channel with adjustable delay between the ventricles. (with the optional module)
- Rapid atrial pacing (overdrive) up to at least 800 per minute to stop atrial flutter, started only by a deliberate guarded action.
- Bedside checks of capture and sensing thresholds by adjusting output and sensitivity while watching the pace and sense indicators and the patient's ECG.
- Runs on ordinary alkaline batteries for at least 5 days and keeps pacing for at least 30 seconds while batteries are changed.
- Runs at least 14 days on one set of batteries, for long post-operative or bridge pacing. (with the optional module)
- One-touch emergency pacing at high output from any state, for staff who must start pacing fast. (with the optional module)
- Audible and visual alarm when a pacing lead is disconnected, broken or short-circuited. (with the optional module)
- Measurement and display of the heart's own P- and R-wave size, with optional automatic sensitivity setting. (with the optional module)
- Hospital biomedical department can check output, rate and sensitivity of every unit with an external pacemaker analyser. (with the optional module)
Not covered: Pacing through skin pads on the chest (transcutaneous pacing) is done by a defibrillator-monitor with a pacing module; this unit needs a pacing wire in or on the heart.; Permanent pacing needs an implantable pacemaker fitted in the cath lab; this unit is only for days to a few weeks of temporary pacing.; Checking or programming implanted pacemakers and measuring lead values at implant needs a pacemaker programmer or pacing system analyser.; Continuous ECG rhythm monitoring and alarms need a patient monitor; the pacemaker only shows pace and sense lights.
Loading the builder...
Standards this specification draws on
- IEC 60601-1 General electrical and mechanical safety of medical electrical equipment.
- IEC 60601-2-31 Particular safety and essential performance of external cardiac pacemakers with internal power source (2020 edition): output accuracy, runaway protection, defibrillation protection and battery depletion behaviour.
- IEC 60601-1-2 Electromagnetic compatibility, so the pacemaker is not inhibited or triggered by phones, electrosurgery and other ICU equipment.
- IEC 60601-1-6 Usability of medical electrical equipment, so settings and locks are hard to misuse.
- IEC 62366-1 Usability engineering process of the manufacturer.
- IEC 62304 Medical device software life-cycle for the pacing timing and control software.
- ISO 14971 Risk management for medical devices.
- ISO 13485 Quality management system of the manufacturer.
- IEC 62353 Recurrent testing and testing after repair of medical electrical equipment.
- DRAP Medical Devices Rules 2017 Registration/enlistment required to import and sell medical devices in Pakistan.
Common mistakes in tenders for this equipment
- Copying one model's brochure: output only in mA (or only in volts), rapid atrial pacing to 1000 ppm, a touch screen or exact sensitivity steps. Accept constant-voltage or constant-current output and use 'covering at least' ranges.
- Buying the pacemaker but not the patient cables, surgical adapters, transvenous catheters or epicardial wires, so the unit sits unused or staff improvise unsafe connections.
- No back-up unit: when the only pacemaker fails or its batteries run out, a pacing-dependent patient has nothing to switch to.
- Accepting refurbished or second-hand units offered as new; temporary pacemakers are widely traded used. Require a brand-new unit, year of manufacture and DRAP registration of the exact model.
- Specifying a proprietary rechargeable battery and charger when ordinary alkaline cells are easier to replace in Pakistan; a dead proprietary battery stops pacing.
- Mixing up this item with transcutaneous pacing in a defibrillator, or with a pacing system analyser used at permanent pacemaker implant.
- Accepting units without testing output, rate and sensitivity on a pacemaker analyser at installation, so there is no baseline for warranty claims.
- Not checking that cable connectors fit the pacing catheters and epicardial wires the hospital already stocks.