Total organic carbon (TOC) analyser for pharmaceutical water and cleaning validation, laboratory and online, with optional conductivity

A TOC analyser oxidises the organic matter in a water sample to carbon dioxide and measures it, giving the total organic carbon content in ppb or mg/L. Pharmaceutical manufacturers and drug testing laboratories use it to release purified water and water for injection under USP <643> / Ph. Eur. 2.2.44 and to check cleaned equipment for residues. Laboratory models test grab samples; online models sit on the water loop and many also measure conductivity.

What this equipment will do

  • Samples: purified water, highly purified water and water for injection from the water system and its user points (USP <643>, Ph. Eur. 2.2.44, JP 2.59).
  • Samples: sterile packaged water, cleaning-validation rinse water and swab extracts, and other aqueous samples up to 50 mg C/L. (with the optional module)
  • Samples: process, raw, waste and environmental water and extracts containing particles or salts (ISO 8245). (with the optional module)
  • Samples: cleaning-validation swabs and solid samples measured directly without extraction. (with the optional module)
  • Test: total organic carbon in mg C/L or ug C/L, with a detection limit of 0.05 mg C/L or lower (USP <643>).
  • Test: system suitability check with sucrose and 1,4-benzoquinone standards, response efficiency 85-115% (USP <643>, Ph. Eur. 2.2.44).
  • Test: continuous or repeated automatic TOC monitoring on the purified water or WFI loop with alarms. (with the optional module)
  • Test: conductivity and temperature of the same sample for Stage 1 water conductivity (USP <645>). (with the optional module)
  • Test: non-purgeable organic carbon (NPOC) with automatic acidification for samples high in carbonate. (with the optional module)
  • Test: total bound nitrogen together with TOC. (with the optional module)
  • Unattended overnight runs of many samples, standards and blanks. (with the optional module)

Not covered: TOC does not identify which compound is present; identifying or quantifying a specific drug or detergent residue needs HPLC or UV-Vis spectrophotometry.; Microbial counts and bacterial endotoxins in water need the microbiology lab (membrane filtration, incubators) and the endotoxin (LAL) reader.; Full USP <645> Stage 2 and 3 conductivity and pH of water need a laboratory conductivity meter and pH meter; nitrates and heavy or elemental impurities need UV-Vis or ICP instruments.

10 international manufacturers checkedMarket checked 2026-1011 standards referenced

Loading the builder...

Standards this specification draws on

  • USP <643> Total Organic Carbon chapter: sets the 0.05 mg C/L detection limit, complete oxidation and the sucrose / 1,4-benzoquinone system suitability test the analyser must pass.
  • Ph. Eur. 2.2.44 European Pharmacopoeia TOC method, harmonised with USP <643> and used through BP monographs for purified water and WFI in Pakistan.
  • JP 2.59 Japanese Pharmacopoeia test for total organic carbon, harmonised with USP <643> and Ph. Eur. 2.2.44.
  • USP <645> Water conductivity chapter: Stage 1 uncompensated conductivity requirements for the optional simultaneous conductivity measurement.
  • USP <1058> Analytical instrument qualification framework (DQ/IQ/OQ/PQ) used to accept the analyser.
  • 21 CFR Part 11 US FDA rule on electronic records and signatures; TOC results used for water and cleaning release are checked against it.
  • EU GMP Annex 11 EU/PIC/S GMP rule for computerised systems, followed in DRAP PIC/S-aligned GMP inspections.
  • ISO 8245:1999 Water quality guideline for TOC and DOC determination, used for research, environmental and process water.
  • IEC 61010-1 Electrical safety standard for laboratory and measurement equipment.
  • IEC 61326-1 EMC requirements for laboratory and measurement equipment.
  • ISO/IEC 17025:2017 Accreditation standard for testing labs; requires calibrated, traceable equipment and controlled records.

Common mistakes in tenders for this equipment

  • Copying one maker's detection principle (for example requiring membrane conductometric or NDIR detection, or a specific turbo or continuous mode name) when the pharmacopoeia accepts any technique that passes the system suitability test.
  • Writing only 'USP <643> compliant' without asking for the 0.05 mg C/L detection limit, a range up to the samples you will really test, and a live system suitability test at OQ.
  • Buying a low-range pure-water analyser for cleaning-validation rinses and swab extracts, or a combustion analyser for ppb-level WFI release without checking its blank, so results go over-range or fail system suitability.
  • Leaving out certified low-TOC vials, system suitability standards, reagents, lamps and carrier gas, so the analyser stands idle or gives false failures from contaminated vials.
  • Ordering an online analyser without sample conditioning, installation on the loop and connection to the plant monitoring system, so it cannot be used on a hot WFI loop.
  • Accepting a 'Part 11 compliant' claim without a demonstration of user roles, audit trail and electronic signatures on the offered software version.
  • Not fixing reagent, standard and spare-part prices and availability for 5-10 years, which lets the single supplier raise consumable prices after installation.
  • Asking for DRAP medical-device registration: a TOC analyser is laboratory / process test equipment, so this demand only confuses bidders.