Ultra-high-performance liquid chromatography (UHPLC) system, 1000 bar or more, with PDA / UV detection

A UHPLC system is a liquid chromatograph built to work at 1000 bar or more with sub-2 um and core-shell columns, giving separations 3-10 times faster than conventional HPLC with less solvent and sharper peaks. Pharmaceutical QC labs with heavy workloads, drug testing laboratories, generic product development groups and research labs buy it for fast assay, impurity and stability methods. It needs low-dispersion hardware, a fast detector and trained staff, and existing pharmacopoeial HPLC methods must be transferred within allowed adjustments or revalidated.

What this equipment will do

  • Samples: dissolved pharmaceutical raw materials and finished products, including content uniformity, dissolution and stability samples, tested against BP/USP/Ph. Eur. monographs (USP <621>, Ph. Eur. 2.2.46).
  • Fast assay and identification: each ingredient separated on sub-2 um or core-shell columns at 1000 bar or more, measured by UV absorbance and identified by retention time.
  • Related substances, impurities and degradation products by gradient methods, with the precision needed for method validation (ICH Q2(R2)).
  • Existing conventional HPLC methods on 4.6 mm columns, or transferred to UHPLC within the allowed adjustments, with matching gradient delay volume (USP <621>, USP <1224>). (with the optional module)
  • Unstable samples kept cooled (4 deg C) in the autosampler during long stability and impurity sequences. (with the optional module)
  • Automatic system suitability results: resolution, tailing, plate count, signal-to-noise and %RSD with pass/fail limits (USP <621>, Ph. Eur. 2.2.46). (with the optional module)
  • Peak purity check and spectral library search from PDA spectra, to show no impurity hides under the main peak. (with the optional module)
  • Naturally fluorescent or derivatised compounds at trace levels, by fluorescence detector. (with the optional module)
  • Compounds with weak or no UV absorbance (e.g. sugars, lipids, some excipients) in gradient methods, by charged aerosol or ELSD detector. (with the optional module)
  • Proteins, peptides and high-salt buffer methods on a metal-free (bio-inert) flow path at 1000 bar. (with the optional module)
  • Unattended method development: automatic screening of several columns and solvents in one sequence. (with the optional module)
  • Audit-trailed, electronically signed results that cannot be edited or deleted, for GMP and DRAP data integrity inspections (21 CFR Part 11, EU GMP Annex 11, WHO TRS 1033 Annex 4). (with the optional module)

Not covered: Residual solvents and other volatile impurities need a gas chromatograph (GC) with headspace sampler.; Elemental impurities and heavy metals (ICH Q3D) need an ICP-MS or ICP-OES.; Molecular mass and structure of unknown impurities need an LC-MS; this pack can only make the system ready for a mass spectrometer to be added later.; Dissolving tablets and capsules for the dissolution test itself needs a dissolution tester, and water content needs a Karl Fischer titrator; the UHPLC only measures the samples they produce.

8 international manufacturers checkedMarket checked 2026-1011 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 61010-1 Basic electrical safety standard for laboratory equipment, including fire risk near flammable solvents.
  • IEC 61326-1 EMC requirements for laboratory equipment so the system neither disturbs nor is disturbed by nearby electronics.
  • USP <621> Defines system suitability and, since harmonisation, the allowed adjustments for moving isocratic and gradient methods to smaller particles and columns.
  • Ph. Eur. 2.2.46 Harmonised European chapter on chromatographic separation used with BP methods in Pakistan, including allowed adjustments.
  • USP <1058> Analytical instrument qualification framework (DQ/IQ/OQ/PQ) used to accept the system.
  • USP <1224> Guidance on transfer of analytical procedures, relevant when moving existing methods onto the UHPLC.
  • ICH Q2(R2) Guideline for validating analytical procedures run on the system, including revalidation of transferred methods.
  • 21 CFR Part 11 US FDA rule on electronic records and signatures; CDS audit trails and e-signatures are checked against it.
  • EU GMP Annex 11 EU/PIC/S GMP rule for computerised systems; applies to the CDS in GMP labs.
  • WHO TRS 1033 Annex 4 WHO guideline on data integrity, relevant for DRAP GMP inspections and WHO prequalification.
  • ISO/IEC 17025:2017 Accreditation standard for testing labs; needs qualified, calibrated equipment and controlled data.

Common mistakes in tenders for this equipment

  • Buying a 1000+ bar UHPLC for a lab that will only run 250 x 4.6 mm, 5 um monograph methods; it costs more and gives no benefit unless methods are transferred.
  • Copying one brand's exact figures (such as a 1300 bar or 18,000 psi limit, a specific delay volume or a 60 mm flow cell) so only that brand complies; state minimum needs with 'not less than' instead.
  • Comparing gradient delay volume, dispersion, noise and carryover figures measured under different conditions; ask bidders to state the test method and verify at OQ.
  • Leaving out UHPLC-rated columns, 0.2 um filters and spare high-pressure fittings, so the system cannot run or columns clog in the first weeks.
  • Moving pharmacopoeial gradient methods to UHPLC without checking the allowed adjustments in USP <621> / Ph. Eur. 2.2.46, then facing audit findings or full revalidation.
  • Using proprietary names for flow cells, mixers, delay-volume features or software modules in the specification instead of measurable performance.
  • Accepting a 'Part 11 compliant' claim without a live demonstration of audit trail, user roles, e-signatures and protection against data deletion.
  • Asking for DRAP medical-device registration; a UHPLC is laboratory equipment, so this demand only confuses bidders.