Specification builder · Tier A

Oil-free laboratory diaphragm vacuum pump for membrane filtration, solvent filtration and degassing, desiccators and SPE manifolds (chemically resistant and vacuum-controller options)

An oil-free diaphragm vacuum pump is the bench vacuum source of a testing lab. It pulls liquids through membrane filters for sterility, microbial limit and particulate tests, filters and degasses HPLC mobile phases, and empties desiccators and SPE manifolds. It is bought by drug testing laboratories, pharma and device QC labs and universities; chemistry labs need the chemically resistant version.

What this equipment will do

  • Sample types: aqueous solutions, rinse and extraction fluids, water samples and dissolved products filtered through 47-50 mm membranes on single funnels or a 3-place manifold (USP <61>, Ph. Eur. 2.6.12, ISO 11737-1).
  • Sample types: HPLC mobile phases and organic solvents (methanol, acetonitrile, ethanol) filtered and degassed under vacuum. (with the optional module)
  • Membrane filtration for the sterility test of injectables on open stainless manifolds (USP <71>, Ph. Eur. 2.6.1). (with the optional module)
  • Membrane filtration for microbial limit (enumeration) tests and water testing (USP <61>, Ph. Eur. 2.6.12).
  • Membrane filtration of injections for the microscopic particulate matter count (USP <788>).
  • Vacuum down to not more than 200 mbar absolute with 15 L/min or more of flow; not more than 100 mbar with 20 L/min or more on the standard pump; not more than 20 mbar on the two-stage pump (measured as in ISO 21360-1).
  • Evacuating vacuum desiccators for drying and storing samples and reference standards. (with the optional module)
  • Running a solid-phase extraction (SPE) manifold for sample clean-up before HPLC or GC. (with the optional module)
  • Holding a set vacuum automatically with digital read-out in mbar, for gentle degassing and repeatable filtration. (with the optional module)
  • Recovering solvent vapour from the pump exhaust. (with the optional module)
  • Aspirating liquid waste and culture supernatants into a 4 L bottle with a hand operator. (with the optional module)
  • Recording the vacuum against time for each run. (with the optional module)

Not covered: It is not a clinical suction machine for patients; that is a medical device (electric suction machine) needing DRAP registration.; It does not run closed-canister sterility tests; those need a sterility test peristaltic pump and canisters (sterility testing pump and manifold).; It does not give the high vacuum needed by freeze dryers, vacuum ovens for residual solvents below a few mbar or mass spectrometers; those need rotary-vane, scroll or turbomolecular pumps.; It does not degas mobile phase online during an HPLC run; that is done by the HPLC in-line degasser (or helium sparging / an ultrasonic bath).

7 international manufacturers checkedMarket checked 2026-1011 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 61010-1 Basic electrical safety standard for laboratory equipment, covering the pump motor, controller and mains wiring.
  • IEC 61326-1 EMC requirements for laboratory equipment, so the motor and controller do not disturb nearby balances and instruments.
  • ISO 21360-1 Standard methods for measuring vacuum-pump performance (ultimate pressure and pumping speed), so datasheet figures are comparable.
  • USP <71> Sterility tests; the membrane filtration method needs a reliable vacuum source for open manifolds.
  • Ph. Eur. 2.6.1 European Pharmacopoeia sterility test (harmonised with USP <71>), used with BP monographs in Pakistan.
  • USP <61> Microbial enumeration tests on non-sterile products, including the membrane filtration method.
  • Ph. Eur. 2.6.12 European Pharmacopoeia microbial enumeration tests (harmonised with USP <61>).
  • USP <788> Particulate matter in injections; the microscopic particle count method filters the sample through a membrane under vacuum.
  • ISO 11737-1 Determination of bioburden on medical devices, often by membrane filtration of extraction fluid.
  • ISO/IEC 17025:2017 Accreditation standard for testing labs; requires calibrated equipment where readings affect results.
  • USP <1058> Analytical instrument qualification framework (simple equipment is qualified by IQ/OQ against its specification).

Common mistakes in tenders for this equipment

  • Writing vacuum as "inches of mercury" or "-0.09 MPa" without saying gauge or absolute and at what altitude, so pumps cannot be compared; specify ultimate vacuum in mbar absolute and flow in L/min at 50 Hz.
  • Copying one brand's exact flow, weight or noise figure so only that brand complies; set minimum vacuum and flow for the number of filtration places you run.
  • Buying a standard rubber-diaphragm pump for HPLC solvent filtration, so the diaphragm fails within months.
  • Forgetting the trap bottle, hydrophobic protective filter and proper vacuum tubing, so filtrate is sucked into the pump in the first week.
  • Asking for an electronic controller, condenser or data logging for simple filtration, which adds cost and removes lower-cost bidders.
  • Not buying spare diaphragms and valves or a parts guarantee, leaving the lab without vacuum for weeks while parts are imported.
  • Asking for DRAP medical-device registration or ISO 13485, or accepting a clinical suction machine in place of a laboratory pump.