Specification builder · Tier A

Single-channel volumetric (large-volume) infusion pump for general, critical care, paediatric and oncology use

A volumetric infusion pump drives fluid from a bag or bottle through an infusion set at a precise programmed rate, typically 1-1200 mL/h, for maintenance fluids, antibiotics, electrolytes, chemotherapy, parenteral nutrition and drug infusions. Hospitals buy them in large numbers for wards, ICU/HDU, OT, emergency, paediatrics and oncology. The key tender decisions are which infusion sets the pump accepts (and what they cost), accuracy with test data, air and occlusion detection, a drug library with dose limits, and battery time for load-shedding.

What this equipment will do

  • Patients: adults and children on wards, ICU/HDU, theatres, emergency and oncology units, receiving maintenance fluids, antibiotics, electrolytes, drug infusions, chemotherapy and parenteral nutrition.
  • Programmed infusion from 1 to 1000 mL/h by rate, volume and time, volume up to 9999 mL, keep-vein-open rate and end-of-infusion alarms, with flow accuracy of +/-5% (IEC 60601-2-24).
  • Works with ordinary gravity IV sets (ISO 8536-4) of at least 3 calibrated brands, 2 of them sold in Pakistan by other suppliers. (with the optional module)
  • Low rates for children and potent drugs: 0.1 to 1200 mL/h with trumpet-curve test data, or 0.01 mL/h steps below 10 mL/h. (with the optional module)
  • Safety alarms: air bubbles of 250 uL or less, blocked line above or below the pump with reduced bolus after release, door open and set misloaded, and free-flow protection (IEC 60601-2-24, IEC 60601-1-8).
  • Finer air protection: adjustable bubble threshold down to 50 uL and alarm on air building up from many small bubbles. (with the optional module)
  • Live line-pressure display against the set limit, and early warning of rising or falling pressure before an occlusion alarm. (with the optional module)
  • Keeps infusing on its own battery during load-shedding and patient transfer: at least 4 hours at 25 mL/h with automatic changeover from mains.
  • Weight-based dose programming (mcg/kg/min, units/h and others) with automatic mL/h conversion, and programmed bolus and line purge. (with the optional module)
  • Chemotherapy and nutrition protocols: ramp up and taper down, intermittent and multi-step programmes, delayed start and secondary (piggyback) infusion. (with the optional module)
  • Drug library with hard and soft dose limits per care area, to stop dosing errors, with a log of limit alerts and overrides for pharmacy review. (with the optional module)
  • Central view of all pumps (drug, rate, volume, alarms) through a dock or wireless network, with data sent to the hospital or ICU information system (HL7) or orders sent from the EMR. (with the optional module)

Not covered: Very low neonatal rates and small volumes of concentrated drugs (sedation, inotropes in newborns, insulin) are given with a syringe pump.; Patient-controlled pain relief (PCA) and home or ambulatory infusion need a PCA pump or an ambulatory infusion pump.; Massive transfusion at rates far above 1200 mL/h with warming needs a rapid infuser or a blood and fluid warmer.; Tube feeding into the stomach or gut needs an enteral feeding pump, never an IV infusion pump.

7 international manufacturers checkedMarket checked 2026-1011 standards referenced

Loading the builder...

Standards this specification draws on

  • IEC 60601-1 General basic safety and essential performance of medical electrical equipment.
  • IEC 60601-1-2 Electromagnetic compatibility, so the pump keeps infusing correctly near diathermy, phones, defibrillators and other equipment.
  • IEC 60601-1-8 Alarm system design: priorities, sounds and visual signals for air-in-line, occlusion, end-of-infusion and battery alarms.
  • IEC 60601-2-24 Particular standard for infusion pumps and controllers: flow accuracy and trumpet-curve testing, air-in-line and occlusion alarms, post-occlusion bolus, free-flow and over-infusion protection.
  • IEC 62366-1 Usability engineering, to reduce programming and set-loading errors.
  • IEC 62304 Software life-cycle for the pump firmware, dose calculations and drug library.
  • ISO 14971 Risk management for medical devices, covering hazards such as free flow, air embolism, over-infusion and missed occlusion.
  • ISO 13485 Quality management system of the manufacturer.
  • ISO 8536-4 Single-use gravity-feed infusion sets; the reference for accepting ordinary sets in open-system pumps and for set quality.
  • IEC 62353 Electrical safety testing at installation, after repair and during preventive maintenance.
  • DRAP Medical Devices Rules 2017 Registration/enlistment required to import and sell medical devices, including infusion pumps and their sets, in Pakistan.

Common mistakes in tenders for this equipment

  • Writing 'infusion sets as recommended by the manufacturer' without a price. Dedicated pump sets can cost USD 3-10 each against well under USD 1 for a gravity set; require the set list and a fixed 3-year set price, and compare 5-year cost, not pump price.
  • Copying one brand's brochure: exact rate ranges such as 0.1-2000 mL/h, a fixed number of occlusion levels or air thresholds, or named software features that only one model has. Use 'not less than' values and accept any equivalent design.
  • Accepting an open-system pump without testing it with the sets the hospital actually buys. Accuracy depends on the tubing; check flow with each named local set brand at acceptance.
  • Quoting headline accuracy only. Ask for the IEC 60601-2-24 start-up graph and trumpet curves at 25 mL/h and 1 mL/h for the offered pump and set.
  • Battery time stated without conditions ('up to 12 hours'). Specify hours at 25 mL/h with a new battery and wireless switched on if offered.
  • Buying pumps with a drug library but no editor licence, no training and no one to build the library, so pumps run without limits and the dose error reduction never protects anyone.
  • Using volumetric pumps for neonatal and very low-rate potent drug infusions where syringe pumps are the safer choice, or buying one type for every unit instead of a planned mix.
  • Accepting DRAP registration of a different model, of the brand only, or of the pump without its dedicated sets; the exact offered pump and its sets must be covered.